ArticleIEEE transactions on bio-medical engineering2021

A Robotically Steerable Guidewire With Forward-Viewing Ultrasound: Development of Technology for Minimally-Invasive Imaging.

Graham C Collins, Achraj Sarma, Zachary L Bercu, Jaydev P Desai, Brooks D Lindsey

Open access · greenAbstract read
In one paragraph

Article in IEEE transactions on bio-medical engineering, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
0.5field-weighted citation impact, top 38% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 19 citations in OpenAlex.

  1. Article
  2. A 1.11 mmIEEE transactions on biomedical circuits and systems · 2025
    Article
  3. Article
  4. High-Frequency, 2-mm-Diameter Forward-Viewing 2-D Array for 3-D Intracoronary Blood Flow Imaging.IEEE transactions on ultrasonics, ferroelectrics, and frequency control · 2024
    Article
  5. Endovascular Detection of Catheter-Thrombus Contact by Vacuum Excitation.IEEE transactions on bio-medical engineering · 2024
    Article
  6. Article
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 1 institution in 1 country.

Graham C Collins
Achraj Sarma
Zachary L Bercu
Jaydev P Desai
Brooks D Lindsey
Georgia Institute of Technology · US

Funding

Ultrasound-guided, Robotically Steerable Guidewire for Endovascular InterventionsR01HL144714 · NHLBI · GEORGIA INSTITUTE OF TECHNOLOGY · PI DESAI, JAYDEV P. · 2019 to 2022
$2.8M
NHLBI NIH HHS R01 HL144714
6 · The paper itself

Abstract

objectiveThe current standard of care for peripheral chronic total occlusions involves the manual routing of a guidewire under fluoroscopy. Despite significant improvements in recent decades, navigation remains clinically challenging with high rates of procedural failure and iatrogenic injury. To address this challenge, we present a proof-of-concept robotic guidewire system with forward-viewing ultrasound imaging to allow visualization and maneuverability through complex vasculature.

methodsA 0.035" guidewire-specific ultrasound transducer with matching layer and acoustic backing was designed, fabricated, and characterized. The effect of guidewire motion on signal decorrelation was assessed with simulations and experimentally, driving the development of a synthetic aperture beamforming approach to form images as the transducer is steered on the robotic guidewire. System performance was evaluated by imaging wire targets in water. Finally, proof-of-concept was demonstrated by imaging an ex vivo artery.

resultsThe designed custom transducer was fabricated with a center frequency of 15.7 MHz, 45.4% fractional bandwidth, and 31 dB SNR. In imaging 20 μm wire targets at a depth of 6 mm, the lateral -6 dB target width was 0.25 ± 0.03 mm. The 3D artery reconstruction allowed visualization of vessel wall structure and lumen.

conclusionInitial proof-of-concept for an ultrasound transducer-tipped steerable guidewire including 3D image formation without an additional sensor to determine guidewire position was demonstrated for a sub-mm system with an integrated ultrasound transducer and a robotically-steered guidewire. SIGNIFICANCE: The developed forward-viewing, robotically-steered guidewire may enable navigation through occluded vascular regions that cannot be crossed with current methods.

Indexed as

TechnologyTransducersEquipment DesignPhantoms, ImagingUltrasonography

Identifiers

PMID33264091
PMCPMC8279262
OpenAlexW3108759020

What Socratic holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the Socratic graph.