ArticleCancers2023
Validation of a Temperature-Feedback Controlled Automated Magnetic Hyperthermia Therapy Device.
Article in Cancers, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 16 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
16 citing papers in PubMed.
- Feasibility of thermal dose painting during magnetic nanoparticle thermal therapy: a computational analysis.International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group · 2026Article
- Verification and validation in canine cadaver of an MPI-integrable RF head coil for magnetic particle hyperthermia.International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group · 2026Article
- Design, verification and validation of a spatially confined magnetic nanoparticle thermal therapy device for conformal thermal dose contouring.International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group · 2026Article
- Verification and Validation of an Autotuning Proportional-Integral-Derivative Controller for Spatially Confined Magnetic Particle Hyperthermia.Journal of medical devices · 2026Article
- A Theoretical Framework for Ligand-Functionalised Magnetic Lipid Nanoparticles in Glioblastoma Therapy.Cancers · 2025Review
- Simulation-Based Design and Machine Learning Optimization of a Novel Liquid Cooling System for Radio Frequency Coils in Magnetic Hyperthermia.Bioengineering (Basel, Switzerland) · 2025Article
- Imaging-guided precision hyperthermia with magnetic nanoparticles.Nature reviews bioengineering · 2025Article
- Iron oxide nanoparticles coated with bioactive materials: a viable theragnostic strategy to improve osteosarcoma treatment.Discover nano · 2025Review
- Article
- Magnetic particle imaging resolution needed for magnetic hyperthermia treatment planning: a sensitivity analysis.Frontiers in thermal engineering · 2025Article
- Numerical Simulation of Thermal Therapy for Melanoma in Mice.Bioengineering (Basel, Switzerland) · 2024Article
- Magnetic Hyperthermia Therapy for High-Grade Glioma: A State-of-the-Art Review.Pharmaceuticals (Basel, Switzerland) · 2024Review
- Advances in magnetic induction hyperthermia.Frontiers in bioengineering and biotechnology · 2024Review
- A Novel Port to Facilitate Magnetic Hyperthermia Therapy for Glioma.Journal of biomechanical engineering · 2024Article
- Novel approaches to targeting gliomas at the leading/cutting edge.Journal of neurosurgery · 2023Review
- Design of a temperature-feedback controlled automated magnetic hyperthermia therapy device.Frontiers in thermal engineering · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
14 authors.
Funding
Abstract
We present in vivo validation of an automated magnetic hyperthermia therapy (MHT) device that uses real-time temperature input measured at the target to control tissue heating. MHT is a thermal therapy that uses heat generated by magnetic materials exposed to an alternating magnetic field. For temperature monitoring, we integrated a commercial fiber optic temperature probe containing four gallium arsenide (GaAs) temperature sensors. The controller device used temperature from the sensors as input to manage power to the magnetic field applicator. We developed a robust, multi-objective, proportional-integral-derivative (PID) algorithm to control the target thermal dose by modulating power delivered to the magnetic field applicator. The magnetic field applicator was a 20 cm diameter Maxwell-type induction coil powered by a 120 kW induction heating power supply operating at 160 kHz. Finite element (FE) simulations were performed to determine values of the PID gain factors prior to verification and validation trials. Ex vivo verification and validation were conducted in gel phantoms and sectioned bovine liver, respectively. In vivo validation of the controller was achieved in a canine research subject following infusion of magnetic nanoparticles (MNPs) into the brain. In all cases, performance matched controller design criteria, while also achieving a thermal dose measured as cumulative equivalent minutes at 43 °C (CEM43) 60 ± 5 min within 30 min.
Indexed as
Identifiers
What Socratic holds
Registered trials