ArticleBMC medical imaging2023

Assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device.

Anssi-Kalle Heikkinen, Valtteri Rissanen, Antti A Aarnisalo, Kristofer Nyman, Saku T Sinkkonen, Juha Koivisto

Open access · goldAbstract read
In one paragraph

Article in BMC medical imaging, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact, top 95% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed, 0 citations in OpenAlex.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 1 institution in 1 country.

Anssi-Kalle HeikkinenDepartment of Otorhinolaryngology-Head and Neck Surgery and Tauno Palva Laboratory, Head and Neck Center, University of Helsinki and Helsinki University Hospital, Helsinki, Finland. anssi-kalle.heikkinen@helsinki.fi.
Valtteri RissanenDepartment of Otorhinolaryngology-Head and Neck Surgery and Tauno Palva Laboratory, Head and Neck Center, University of Helsinki and Helsinki University Hospital, Helsinki, Finland.
Antti A AarnisaloDepartment of Otorhinolaryngology-Head and Neck Surgery and Tauno Palva Laboratory, Head and Neck Center, University of Helsinki and Helsinki University Hospital, Helsinki, Finland.
Kristofer NymanRadiology, HUS Diagnostic Center, University of Helsinki and Helsinki University Hospital, Helsinki, Finland.
Saku T SinkkonenDepartment of Otorhinolaryngology-Head and Neck Surgery and Tauno Palva Laboratory, Head and Neck Center, University of Helsinki and Helsinki University Hospital, Helsinki, Finland.
Juha KoivistoDepartment of Physics, University of Helsinki, Helsinki, Finland.
University of Helsinki · FI

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundMulti slice computed tomography (MSCT) is the most common used method in middle ear imaging. However, MSCT lacks the ability to distinguish the ossicular chain microstructures in detail resulting in poorer diagnostic outcomes. Novel cone beam computed tomography (CBCT) devices' image resolution is, on the other hand, better than MSCT resolution. The aim of this study was to optimize imaging parameters of a novel full body CBCT device to obtain optimal contrast to noise ratio (CNR) with low effective dose, and to optimize its clinical usability.

methodsImaging of five anonymous excised human cadaver temporal bones, the acquisition of the effective doses and the CNR measurements were performed for images acquired on using Planmed XFI® full body CBCT device (Planmed Oy, Helsinki, Finland) with a voxel size of 75 µm. All images acquired from the specimens using 10 different imaging protocols varying from their tube current exposure time product (mAs) and tube voltage (kVp) were analyzed for eight anatomical landmarks and evaluated by three evaluators.

resultsWith the exception of protocol with 90 kVp 100 mAs, all other protocols used are competent to image the finest structures. With a moderate effective dose (86.5 µSv), protocol with 90 kV 450 mAs was chosen the best protocol used in this study. A significant correlation between CNR and clinical image quality of the protocols was observed in linear regression model. Using the optimized imaging parameters, we were able to distinguish even the most delicate middle ear structures in 2D images and produce accurate 3D reconstructions.

conclusionsIn this ex vivo experiment, the new Planmed XFI® full body CBCT device produced excellent 2D resolution and easily created 3D reconstructions in middle ear imaging with moderate effective doses. This device would be suitable for middle ear diagnostics and for e.g., preoperative planning. Furthermore, the results of this study can be used to optimize the effective dose by selecting appropriate exposure parameters depending on the diagnostic task.

Indexed as

Cone-Beam Computed TomographyTomography, X-Ray ComputedEar, MiddleHumansPhantoms, ImagingRadiation Dosage3D reconstructionCNRCone beam computed tomographyContrast to noise ratioFull body CBCTMiddle ear imagingSubjective image quality