ArticleEJNMMI physics2023
Gamma camera-specific reference standards for radioactive iodine uptake measurements.
Article in EJNMMI physics, 2023. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 1 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 3 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundCurrent guidelines of the radioiodine uptake (RAIU) test allow the use of different equipment, isotopes, activity and region-of-interest (ROI). We evaluated presence and extent of these differences in clinical practice and evaluated the effect of some of these variations on RAIU outcomes. Also, gamma camera-specific reference standards were calculated and retrospectively compared with measurements obtained during clinical RAIU tests. MATERIALS AND
methodsFirst, questionnaires were sent to Dutch nuclear medicine departments requesting information about equipment usage, isotope, isotope formulation, activity and measurement techniques. Secondly, a neck phantom containing a range of activities in capsule or water-dissolved formulation was scanned. Counts were measured using automatic ROI, square box ROI or all counts in the image. Thirdly, clinical RAIU data were collected during 2015-2018 using three different gamma cameras. Reference standards for each scanner were calculated using regression analysis between reference activity and measured counts. Uptake measurements using this gamma camera-specific reference standard were compared with original measurements.
resultsThe survey demonstrated significant differences in isotope, isotope formulation, activity, use of neck phantoms, frequency and duration of reference measurements, distance to collimator, use of background measurements and ROI delineation. The phantom study demonstrated higher counts for the water-dissolved formulation than capsules using both automatic and square box ROI. Also, higher counts were found using a square box ROI than an automatic ROI. The retrospective study showed feasibility of RAIU calculations using camera-specific reference standards and good correlation with the original RAIU measurements.
conclusionsThis study demonstrated considerable technical variation in RAIU measurement in clinical practice. The phantom study demonstrated that these differences could result in differences in count measurements, potentially resulting in different dose calculations for radioactive iodine therapy. Retrospective data suggest that camera-specific reference standards may be used instead of individual reference measurements using separate activity sources, which may thus eliminate some sources of variation.