Evidence map›Paper›PMID 39014052›Full record

ArticleAnnals of biomedical engineering2024

HeartMate 3 for Heart Failure with Preserved Ejection Fraction: In Vitro Hemodynamic Evaluation and Anatomical Fitting.

Nina Langer, Andrew F Stephens, Michael Šeman, David McGiffin, David M Kaye, Shaun D Gregory

Abstract read
In one paragraph

Article in Annals of biomedical engineering, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Article
  3. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Nina LangerCardio-Respiratory Engineering and Technology Laboratory (CREATElab), Department of Mechanical and Aerospace Engineering, Monash University, Melbourne, VIC, Australia. nina.langer@monash.edu.ORCID http://orcid.org/0009-0005-5175-929X
Andrew F StephensCardio-Respiratory Engineering and Technology Laboratory (CREATElab), Department of Mechanical and Aerospace Engineering, Monash University, Melbourne, VIC, Australia.ORCID http://orcid.org/0000-0002-2271-750X
Michael ŠemanCardio-Respiratory Engineering and Technology Laboratory (CREATElab), Department of Mechanical and Aerospace Engineering, Monash University, Melbourne, VIC, Australia.ORCID http://orcid.org/0000-0003-2555-3325
David McGiffinDepartment of Cardiothoracic Surgery, The Alfred, Melbourne, VIC, Australia.ORCID http://orcid.org/0000-0003-4894-2518
David M KayeThe Department of Cardiology, The Alfred Hospital, Melbourne, VIC, Australia.ORCID http://orcid.org/0000-0002-5188-5289
Shaun D GregoryCardio-Respiratory Engineering and Technology Laboratory (CREATElab), Department of Mechanical and Aerospace Engineering, Monash University, Melbourne, VIC, Australia.ORCID http://orcid.org/0000-0002-5448-1917

Funding

National Health and Medical Research Council 2016995National Heart Foundation of Australia 106675
6 · The paper itself

Abstract

Heart failure with preserved ejection fraction (HFpEF) constitutes approximately 50% of heart failure (HF) cases, and encompasses different phenotypes. Among these, most patients with HFpEF exhibit structural heart changes, often with smaller left ventricular cavities, which pose challenges for utilizing ventricular assist devices (VADs). A left atrial to aortic (LA-Ao) VAD configuration could address these challenges, potentially enhancing patient quality of life by lowering elevated mean left atrial pressure (MLAP). This study assessed the anatomical compatibility and left atrial unloading capacity using a simulated VAD-supported HFpEF patient. A HeartMate3-supported HFpEF patient in an LA-Ao configuration was simulated using a cardiovascular simulator. Hemodynamic parameters were recorded during rest and exercise at seven pump flow rates. Computed tomography scans of 14 HFpEF (NYHA II-III) and six heart failure with reduced ejection fraction patients were analysed for anatomical comparisons. HFpEF models were independently assessed for virtual anatomical fit with the HM3 in the LA-Ao configuration. Baseline MLAP was reduced from 15 to 11 mmHg with the addition of 1 L/min HM3 support in the rest condition. In an exercise simulation, 6 L/min of HM3 support was required to reduce the MLAP from 29 to 16 mmHg. The HM3 successfully accommodated six HFpEF patients without causing interference with other cardiac structures, whereas it caused impingement ranging from 4 to 14 mm in the remaining patients. This study demonstrated that the HM3 in an LA-Ao configuration may be suitable for unloading the left atrium and relieving pulmonary congestion in some HFpEF patients where size-related limitations can be addressed through pre-surgical anatomical fit analysis.

Indexed as

Heart-Assist DevicesHeart FailureHemodynamicsModels, CardiovascularStroke VolumeAgedFemaleHumansMaleMiddle AgedAcute heart failureAnatomical fittingCardiovascular simulatorHFpEFLeft ventricular assist deviceMechanical circulatory support