Evidence mapPaperPMID 39326038Full record

ArticleJMIR human factors2024

Triage Accuracy and the Safety of User-Initiated Symptom Assessment With an Electronic Symptom Checker in a Real-Life Setting: Instrument Validation Study.

Ville Liu, Minna Kaila, Tuomas Koskela

Abstract readValidation Study
In one paragraph

Article in JMIR human factors, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Trial
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Ville LiuFaculty of Medicine, University of Helsinki, Helsinki, Finland.ORCID 0000-0002-8441-3480
Minna KailaPublic Health Medicine, Faculty of Medicine, University of Helsinki, Helsinki, Finland.ORCID 0000-0002-9645-4925
Tuomas KoskelaDepartment of General Practice, Faculty of Medicine and Health Technology, Tampere University, Tampere, Finland.ORCID 0000-0002-1518-4944

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundPrevious studies have evaluated the accuracy of the diagnostics of electronic symptom checkers (ESCs) and triage using clinical case vignettes. National Omaolo digital services (Omaolo) in Finland consist of an ESC for various symptoms. Omaolo is a medical device with a Conformité Européenne marking (risk class: IIa), based on Duodecim Clinical Decision Support, EBMEDS.

objectiveThis study investigates how well triage performed by the ESC nurse triage within the chief symptom list available in Omaolo (anal region symptoms, cough, diarrhea, discharge from the eye or watery or reddish eye, headache, heartburn, knee symptom or injury, lower back pain or injury, oral health, painful or blocked ear, respiratory tract infection, sexually transmitted disease, shoulder pain or stiffness or injury, sore throat or throat symptom, and urinary tract infection). In addition, the accuracy, specificity, sensitivity, and safety of the Omaolo ESC were assessed.

methodsThis is a clinical validation study in a real-life setting performed at multiple primary health care (PHC) centers across Finland. The included units were of the walk-in model of primary care, where no previous phone call or contact was required. Upon arriving at the PHC center, users (patients) answered the ESC questions and received a triage recommendation; a nurse then assessed their triage. Findings on 877 patients were analyzed by matching the ESC recommendations with triage by the triage nurse.

resultsSafe assessments by the ESC accounted for 97.6% (856/877; 95% CI 95.6%-98.0%) of all assessments made. The mean of the exact match for all symptom assessments was 53.7% (471/877; 95% CI 49.2%-55.9%). The mean value of the exact match or overly conservative but suitable for all (ESC's assessment was 1 triage level higher than the nurse's triage) symptom assessments was 66.6% (584/877; 95% CI 63.4%-69.7%). When the nurse concluded that urgent treatment was needed, the ESC's exactly matched accuracy was 70.9% (244/344; 95% CI 65.8%-75.7%). Sensitivity for the Omaolo ESC was 62.6% and specificity of 69.2%. A total of 21 critical assessments were identified for further analysis: there was no indication of compromised patient safety.

conclusionsThe primary objectives of this study were to evaluate the safety and to explore the accuracy, specificity, and sensitivity of the Omaolo ESC. The results indicate that the ESC is safe in a real-life setting when appraised with assessments conducted by triage nurses. Furthermore, the Omaolo ESC exhibits the potential to guide patients to appropriate triage destinations effectively, helping them to receive timely and suitable care. INTERNATIONAL REGISTERED REPORT IDENTIFIER (IRRID): RR2-10.2196/41423.

Indexed as

Symptom AssessmentTriageAdultAgedFemaleFinlandHumansMaleMiddle AgedPrimary Health CareeHealthemergency department triagehealth serviceshealth services accessibilityhealth services researchnurse triageprimary careprimary health careremote consultationsymptom assessmenttelemedicinetriageurgent care