ArticleBMC ophthalmology2024
Repeatability and agreement of two swept-source optical coherence tomographers and scheimpflug imaging for measurements of corneal curvature.
Article in BMC ophthalmology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed.
- Interdevice Agreement of Keratometry, Astigmatism Vectors, and Ocular Biometry in Cataract Candidates: SS-OCT (Argos) vs. OLCI (Aladdin) vs. Scheimpflug-Placido (Sirius).Bioengineering (Basel, Switzerland) · 2026Article
- The Effect of Tear Film Stability on Variability of Posterior Corneal Curvature Readings with a Swept-Source Optical Biometer in Cataract Patients.Clinical ophthalmology (Auckland, N.Z.) · 2026Article
- Repeatability of Corneal Astigmatism and Equivalent Power with the MS-39 Tomographer Derived from Model Surface Fitting in a Cataractous Population.Sensors (Basel, Switzerland) · 2025Article
- Crystalline lens power and astigmatism calculation in myopia by swept-source optical coherence tomography: A pilot study.Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2025Article
- Repeatability of corneal curvature with the MS-39 anterior segment tomographer in a cataractous population.PloS one · 2025Article
- Repeatability of corneal measurements from the Casia2 anterior segment tomographer in a cataractous population.PloS one · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo evaluate the repeatability and agreement of two swept-source optical coherence tomographers and Scheimpflug imaging for corneal curvature in healthy subjects to obtain data on the clinical application of a new device.
methodsThis prospective study was conducted in January and February 2021 with 100 healthy subjects at the Eye Hospital of Wenzhou Medical University. Simulated keratometry (Sim-K), posterior keratometry (Kp), total corneal power (TCP), and total corneal astigmatism (TCA) were measured with CASIA2, Anterion, and Pentacam. Within-subject standard deviation (Sw), repeatability coefficient (RC), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were used to assess inter-device repeatability. Bland-Altman analysis was performed to determine inter-device agreement.
resultsThe three devices showed good repeatability for Sim-K, Kp, and TCP with all ICC > 0.980. Pentacam showed the highest repeatability (ICC ≥ 0.993; ICC ≥ 0.993) while the CASIA2 demonstrated the lowest repeatability (ICC: ≥ 0.986; ICC: ≥ 0.985) for Sim-K and TCP. Anterion and CASIA2 revealed better repeatability (ICC ≥ 0.998; ICC ≥ 0.981) for Kp than Pentacam (ICC ≥ 0.980). Pentacam and Anterion showed good repeatability for TCA (ICC: 0.935 and 0.916), whereas the CASIA2 showed moderate repeatability (ICC: 0.836). Three instruments demonstrated good agreement with the maximum absolute 95% Limits of agreement (LoA) of 1.00 D for Sim-K, Kp, and TCP. Wide LoA were found for TCA with the maximum absolute 95% LoA ≥ 0.66 D between the three devices.
conclusionsIn healthy subjects, the three devices (Pentacam, Anterion and CASIA2) displayed comparable repeatability and accuracy for SimK, Kp, and TCP, and could be used interchangeably for these parameters. However, TCA measured by the three devices was not interchangeable.
trial registrationChinese Clinical Trial Registry Center (10/10/2020, ChiCTR2000038959).