Evidence map›Paper›PMID 39474584›Full record

ArticleGalen medical journal2024

A New Remote Monitoring System: Evaluation of the Efficiency and Accuracy of the Smart Emergency Medical System-Health Internet of Things Device.

Mohammad Taghi Hedayati Goudarzi, Hadi Zare Marzouni, Fazel Tarkhan, Ali Bijani, Mehdi Babagoli, Amirhossein Shadifar, Javad Abbas Alipour

Abstract read
In one paragraph

Article in Galen medical journal, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Mohammad Taghi Hedayati GoudarziCardiology Department, Rohani Hospital, School of Medicine, Babol University of Medical Sciences, Babol, Iran.
Hadi Zare MarzouniQaen Faculty of Medical Sciences, Birjand University of Medical Sciences, Birjand, Iran.
Fazel TarkhanBiomedical and Microbial Advanced Technologies Research Center, Health Research Institute, Babol University of Medical Sciences, Babol, Iran.
Ali BijaniDepartment of Epidemiology, Social Determinants of Health Research Center, Health Research Institute, Babol University of Medical Sciences, Babol, Iran.
Mehdi BabagoliIT Engineering Group, Department of Industrial Engineering, K.N. Toosi University of Technology, Tehran, Iran.
Amirhossein ShadifarDepartment of Electrical and Computer Engineering, Faculty of Electrical Engineering, Babol Noshirvani University of Technology, Babol, Iran.
Javad Abbas AlipourClinical Research Development Unit of Rouhani Hospital, Babol University of Medical Sciences, Babol, Iran.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe remote medical monitoring system can facilitate monitoring patients with cardiac arrhythmia, and consequently, reduce mortality and complications in individuals requiring emergency interventions. Hence, it is necessary to evaluate new telemedicine devices and compare them with standard devices. Therefore, this study aimed to evaluate and compare the new remote monitoring system, Smart Emergency Medical System-Health Internet of Things (SEMS-HIOT) developed by the Health Technology Development Centre of Babol University of Medical Sciences on patients with different cardiac arrhythmias and compare it with the standard device. MATERIALS AND

methodsIn this case-control study, 60 patients were divided into the six most common arrhythmia groups (n=10 per each group and equal gender) as atrial fibrillation, ventricular tachycardia, paroxysmal supraventricular tachycardia, premature ventricular contractions, atrial tachycardia, and premature atrial contractions. Also, 20 healthy individuals (including 10 men and 10 women) without any arrhythmia (normal rhythm) were considered as the control group. Three similar SEMS-HIOT devices were used as test devices and a standard cardiac monitoring device as the control device. The clinical parameters, including heart rate, pulse rate, oxygen saturation, body temperature, and cardiac electrical activity via electrocardiogram (ECG) lead-II were recorded.

resultsFindings showed that the performance of the SEMS-HIOT test device was similar and in the same range for all indices in each group and there were no significant differences compared to the performance of the control device (P0.05). Also, the ECG records measured with SEMS-HIOT and standard device indicate no significant differences (P0.05).

conclusionOur study showed that the cardiac indices as well as ECG findings, which were measured with SEMS-HIOT and common standard devices confirmed the accuracy and reliability of the new telematics device for monitoring patients with cardiac diseases.

Indexed as

ArrhythmiasCardiovascular DiseaseTelehealthTelemedicine

Identifiers

PMID39474584
PMCPMC11521570

What Socratic holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the Socratic graph.