ArticleJournal of biomedical optics2024
Development and characterization of silicone-based tissue phantoms for pulse oximeter performance testing.
Article in Journal of biomedical optics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed, 1 synthesis or guideline pooled it.
- Unifying the spectrum: a framework for the meta-analysis of human skin tone scales.Journal of biomedical optics · 2026Pooled it
- Quantitative evaluation of skin pigmentation effects on photoplethysmography using vascular finger phantoms and Monte Carlo simulation.Journal of biomedical optics · 2026Article
- Enhancing skin tone representation in optical vascular phantoms.Journal of biomedical optics · 2026Article
- Optical phantoms for the characterization and validation of imaging and spectroscopy: a review.Biophotonics discovery · 2026Review
- Optical finger phantom with realistic optical properties.Biomedical optics express · 2025Article
- Evaluation of skin pigmentation effect on photoplethysmography signals using a vascular finger phantom with tunable optical and mechanical properties.Journal of biomedical optics · 2025Article
- A Cost-Effective 3D-Printed Conductive Phantom for EEG Sensing System Validation: Development, Performance Evaluation, and Comparison with State-of-the-Art Technologies.Sensors (Basel, Switzerland) · 2025Article
- A Cost-Effective Method for the Spectral Calibration of Photoplethysmography Pulses: The Optimal Wavelengths for Heart Rate Monitoring.Sensors (Basel, Switzerland) · 2025Article
- Pulse Oximetry: 50 Years of Inventions & Discoveries in Biomedical Optics.Journal of biomedical optics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Significance: Pulse oximeter measurements are commonly relied upon for managing patient care and thus often require human testing before they can be legally marketed. Recent clinical studies have also identified disparities in their measurement of blood oxygen saturation by race or skin pigmentation. Aim: The development of a reliable bench-top performance test method based on tissue-simulating phantoms has the potential to facilitate pre-market assessment and the development of more accurate and equitable devices. To generate phantoms capable of mimicking physical mechanisms and providing realistic results, customized tissue-mimicking materials (TMMs) are needed. Approach: We focused on the development of channelized finger phantoms based on flexible silicone elastomers and their implementation in a pulsatile pressurized fluid network. Candidate TMMs were formulated to achieve a range of biologically relevant mechanical and optical properties by modifying components and curing protocols. Results: Our final optimized TMM had a Shore OO hardness of 32 and an elastic modulus of 130 kPa. TMM samples with sub-millimeter diameter channels exhibited compliance-increase in channel diameter with internal fluid pressure, as measured by optical coherence tomography-that was linearly dependent on internal pressure. Phantoms implemented in the pressurized network with an absorber-doped fluid and measured by a photoplethysmographic (PPG) sensor displayed tunable modulation levels ranging from 0.6% to 18.1% at 940 nm. Finally, we demonstrated that the system could be used to generate measurements in several clinical pulse oximeters and variations in PPG waveform could be produced by varying the simulated epidermal melanin content. Conclusions: Overall, we provide significant insights into potential best practices for creating silicone-based tissue phantom tools for pulse oximetry performance testing.