ArticleLangenbeck's archives of surgery2025
Accuracy and feasibility of continuous glucose monitoring system in pancreatectomy patients.
Article in Langenbeck's archives of surgery, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposePancreatectomy patients often experience challenging fluctuations in blood glucose levels; therefore, they require a reliable monitoring system. This study aimed to determine the accuracy and acceptability of a continuous glucose monitoring (CGM) system compared with the intermittent capillary glucose test in patients who have undergone pancreatectomy.
methodsThirty non-diabetic pancreatectomy patients participated. We used the FreeStyle Libre Flash Glucose Monitoring System (Abbott Diabetes Care) for continuous interstitial glucose monitoring. Capillary reference glucose levels were checked four times daily. Accuracy was checked using the Clarke Error Grid.
resultsThe mean age of the participants was 56.8 ± 12.0 years, of whom 61.3% underwent pancreaticoduodenectomy and 38.7% underwent distal pancreatectomy. Three patients developed pancreatogenic diabetes after pancreatectomy. The clinical accuracy of continuous glucose monitoring compared with capillary glucose was 43.9% in Zone A (clinically accurate zone) and 99.8% in Zone A + B (low risk of error) of the Clarke Error Grid. No device-related adverse events were reported. Patients rated favorable user acceptability on the questionnaire.
conclusionThis pilot study demonstrated that the CGM device is accurate and safe for patients who underwent pancreatectomy, with favorable user acceptability. Despite these challenges, the study proposes that the CGM device is beneficial for monitoring glucose levels after discharge in patients with impaired glucose levels following pancreatectomy.