Evidence map›Paper›PMID 39927495›Full record

Observational studyJournal of hypertension2025

Evaluation of Aktiia cuffless blood pressure monitor across 24-h, daytime, and night-time measurements versus ambulatory monitoring: a prospective, single-centre observational study.

Tiago P Almeida, David Perruchoud, Jérémy Alexandre, Pascale Vermare, Josep Sola, Jay Shah, Luisa Marques, Cyril Pellaton

Registry-linked trialAbstract readObservational StudyComparative Study
In one paragraph

Observational study in Journal of hypertension, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT04548986 (Single-center Prospective Study to Validate the Performance of the Aktiia Optical Blood Pressure Monitoring), which is not on this map. Cited by 11 papers.

0numbers the graph read from it
0cells of the map it votes in
11citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT04548986 nacompletednot on this map

Single-center Prospective Study to Validate the Performance of the Aktiia Optical Blood Pressure Monitoring (OBPM) Device at the Wrist Against Ambulatory Blood Pressure Monitoring (ABPM)

TypeinterventionalSponsorCyril PellatonRan2020 to 2022Enrolled63ConditionsBlood Pressure, HypertensionArmsAktiia.product
3 · Its place in the literature

Who cites it

11 citing papers in PubMed.

  1. Article
  2. Article
  3. Disparities in Blood Pressure Measurement.Current hypertension reports · 2026
    Review
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Review
  11. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Tiago P AlmeidaAktiia SA.
David PerruchoudAktiia SA.
Jérémy AlexandreAktiia SA.
Pascale VermareAktiia SA.
Josep SolaAktiia SA.
Jay ShahAktiia SA.
Luisa MarquesDivision of Cardiology, Réseau Hospitalier Neuchâtelois (RHNe), Neuchâtel, Switzerland.
Cyril PellatonDivision of Cardiology, Réseau Hospitalier Neuchâtelois (RHNe), Neuchâtel, Switzerland.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesCuffless blood pressure (BP) monitors represent an innovative approach to BP assessment, overcoming the traditional episodic nature of readings obtained from cuff-based, oscillometric modalities. Nonetheless, the comparison of these cuffless devices against their cuff-based counterparts remain debated.

methodsThis study evaluated BP readings from a cuffless, continual BP monitor (Aktiia monitor, Neuchâtel, Switzerland) and a standard 24-h ambulatory BP monitor (ABPM; Dyasis 3, Novacor, Paris, France) in 54 patients undergoing a 12-week cardiac rehabilitation program in Neuchâtel, Switzerland. Comparisons encompassed 24-h, daytime (9 a.m.-9 p.m.), and night-time (11 p.m.-7 a.m.) systolic and diastolic BP (SBP, DBP) using a 7-day average from Aktiia overlapping a 1-day average from ABPM. Nocturnal dip analysis was performed with 1-day average from Aktiia coinciding with ABPM measurements.

resultsNo significant differences emerged between the Aktiia and ABPM monitors for 24-h and daytime SBP (24-h: 2.6 ± 12.3 [-0.2, 5.4] mmHg, r  = 0.57, P  = 0.06; daytime: 1.2 ± 12.4 [-1.6, 4.0] mmHg, r  = 0.60, P  = 0.38). Night-time SBP showed more evident differences (12.5 ± 14.4 [9.3, 15.8] mmHg, r  = 0.39, P  < 0.001). DBP were within clinical range across 24-h and daytime (24-h: -2.9 ± 7.9 [-4.7, -1.1] mmHg, r  = 0.63, P  = 0.002; daytime: -3.1 ± 8.2 [-5.0, -1.3] mmHg, r  = 0.64, P  = 0.001), and showed significant differences for night-time (4.1 ± 8.5 [2.2, 6.0] mmHg, r  = 0.57, P  < 0.001). The concordance rate with ABPM's nocturnal dip was 79% for 1-day average from Aktiia.

conclusionsThese results demonstrate that the Aktiia monitor yields BP measurements consistent with an ABPM monitor despite some systematic differences during night-time readings. The Aktiia monitor holds potential for future use in continual BP monitoring. CLINICAL

trial registrationNCT04548986. URL: https://clinicaltrials.gov/study/NCT04548986?tab=results.

Indexed as

Blood Pressure Monitoring, AmbulatoryBlood Pressure MonitorsAgedBlood PressureBlood Pressure DeterminationCircadian RhythmFemaleHumansMaleMiddle AgedProspective Studies

Identifiers

PMID39927495
PMCPMC11872257

What Socratic holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390