Quantifying multi-institutional ADC measurement variability of 1.5 T MR-Linacs: A phantom and in vivo study.
Article in Medical physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed.
- Quantitative magnetic resonance imaging for accurate delineation of radiotherapy: a narrative review.Translational cancer research · 2026Review
- A functionally guided fusion Vision Transformer for predicting IDH status in gliomas: a multicenter study with external validation and incomplete multimodal evaluation.Radiologie (Heidelberg, Germany) · 2026Article
- Adaptation of an arterial spin labeling sequence for the assessment of brain perfusion on a 0.35 T MR-Linac.Magma (New York, N.Y.) · 2026Article
- Quantifying multi-institutional ADC measurement variability of 1.5 T MR-Linacs: A phantom and in vivo study.Medical physics · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
12 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundDiffusion-weighted imaging (DWI), a quantitative magnetic resonance imaging (qMRI) technique, has the potential to aid in disease characterization and treatment response monitoring. MR-Linacs (MRLs) enable simultaneous DWI acquisitions during radiotherapy, uniquely aiding in the collection of large-scale datasets for imaging biomarkers, such as the DWI-derived apparent diffusion coefficient (ADC), without additional patient burden. However, the limited data reporting on variability in MRL scanner performance characteristics, and a lack of established clinical trial quality assurance (QA) procedures, are barriers to this route for biomarker validation. PURPOSE: This study aims to quantify the accuracy, intra-scanner repeatability, and inter-scanner reproducibility of ADC measurements across three MRLs in Australia in both a phantom and in vivo. These measurements will inform the feasibility of carrying out prospective multi-center studies in Australia investigating ADC as a biomarker and form a core set of QA procedures and baselines to assess biomarker and sequence suitability.
methodsAn isotropic diffusion phantom (at 0°C) and one healthy volunteer were scanned on three Unity MRLs (Elekta AB, Stockholm, Sweden). Standardized (QIBA Diffusion Profile) and anatomy-specific DWI sequences, including sequences recommended by the MR-Linac Consortium Imaging Biomarker Working Group, were used to image the phantom and volunteer. ADC maps generated using the MRL scanner software (inline ADC) and diffusion-weighted (b-value) images were exported from the scanner console. The latter was used to generate ADC maps using commercial software (offline ADC) for a separate comparative analysis. Performance metrics were computed for each sequence, including a coefficient of variation to assess between-session intra-scanner repeatability (CV
resultsPhantom-based measurements for the standardized QIBA sequence had intra- and inter-scanner CV and bias well within recommended guideline (QIBA Diffusion Profile) tolerance limits of 2.2% and ±3.6%, respectively. All anatomy-specific phantom DWI sequences were also within these tolerances, except for the cervix sequence at one site which showed an average intra-scanner bias of +4.5%. Both accuracy and reproducibility for all sequences were worse for lower diffusivity vials measured in the phantom. Additionally, inline and offline ADC maps had high similarity with average percent differences of +0.2%. Volunteer-based results had worse reproducibility, with the average inter-scanner CV for the brain and pancreas sequences within 9.0%, however, reaching up to 27.1% for pelvis and abdomen sequences.
conclusionsThis study demonstrated accuracy, intra-scanner repeatability, and inter-scanner reproducibility comparable to metrics reported in the literature, using both the phantom and volunteer datasets. The cervix sequence had the largest variability in both phantom and volunteer results and was recommended for further investigation. This study suggests that qMRI techniques utilizing DWI could be a viable option for future multi-centered patient-based studies utilizing Australian MRLs, with phantom-based quality assurance recommended alongside patient imaging.