ArticleFluids and barriers of the CNS2025
VIEshunt: towards a ventricular intelligent and electromechanical shunt for hydrocephalus therapy.
Article in Fluids and barriers of the CNS, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- VIEshunt: towards a ventricular intelligent and electromechanical shunt for hydrocephalus therapy.Fluids and barriers of the CNS · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
17 authors.
Funding
Abstract
backgroundShunt systems for hydrocephalus therapy are commonly based on passive mechanical pressure valves that are driven by the intracranial, intra-abdominal, and hydrostatic pressure. The differential pressure acting on the valve determines the drainage rate of cerebrospinal fluid (CSF) but is not a gauge of the physiological condition of the patient. Internal and external influences can cause over- or underdrainage and lead to pathological levels of intracranial pressure (ICP).
methodsThe first prototype of a ventricular intelligent and electromechanical shunt (VIEshunt) is developed, tested, and compared with previous efforts towards the development of a smart shunt. Its key components are a micro pump, a flow meter, a pressure sensor, an inertial measurement unit, a wireless communication interface, and a microcontroller. The VIEshunt prototype was tested in vitro using a hardware-in-the-loop (HiL) test bench that runs real-time patient simulations involving changes in intracranial and intra-abdominal pressure, as well as changes in posture ranging between supine and upright position. The prototype was subsequently tested in an in vivo pilot study based on an acute ovine animal model (n=1) with infusions of artificial CSF.
resultsDuring 24 h in vitro testing, the prototype detected the simulated posture changes of the patient and automatically adapted the controller reference. The posture-specific ICP references of 12 mmHg for supine and -3 mmHg for upright position were tracked without offset, thus preventing adverse over- and underdrainage during the investigated HiL test scenario. During acute in vivo testing, the prototype first regulated the mean ICP of a sheep from 22 mmHg down to 20 mmHg. Each of the three subsequent intraventricular bolus infusions of 1 mL saline solution increased mean ICP by approximately 11 mmHg. While natural absorption alone decreased ICP by only 5 mmHg within 9 min, the prototype was able to regulate ICP back to the pre-bolus pressure value within 5 min.
conclusionThe developed VIEshunt prototype is capable of posture-dependent ICP regulation and CSF drainage control. Smart shunt systems based on VIEshunt could improve patient monitoring and enable optimal physiologic therapy by adapting to the individual patient. To derive statistically relevant conclusions for the performance of VIEshunt, future work will focus on the development of a next generation prototype for use in pre-clinical studies.
Indexed as
Identifiers
What Socratic holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the Socratic graph.