Real-time blood gas management: evaluating quantum perfusion system's accuracy against a standard blood gas analysis in CPB.
Article in Journal of cardiothoracic surgery, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundContinuous blood gas monitoring (CBGM) during cardiopulmonary bypass (CPB) is essential for maintaining optimal patient outcomes, enabling rapid responses to critical fluctuations in blood gas parameters. This non-inferiority study evaluates the Quantum Perfusion System by Spectrum Medical, which features continuous online blood gas monitoring through Quantum workstation (QWS) and Quantum ventilation module (QVM) without the use of cuvettes, against the standard blood gas analysis (BGA) analyzer to assess real-time clinical accuracy.
methodsThis retrospective study included a sample of 40 patients, monitored continuously with the QPS and compared at intervals against standard BGA measurements. The patients undergoing on elective CPB procedures, specifically for coronary artery bypass grafting (CABG), mitral valve replacement (MVR), and aortic valve replacement (AVR).
resultsPre-alignment deviations for all parameters were within CLIA thresholds, confirming baseline reliability. For hemoglobin, the pre-alignment deviation was 1.9%, which decreased to 0.7% post-alignment, both within the CLIA threshold of ± 5%, with a Bland-Altman mean difference of 0.0988 g/dL (limits: 0.0963 to 0.1012 g/dL). Hematocrit showed a pre-alignment deviation of 2.1%, reduced to 0.2% post-alignment, both within the CLIA threshold of ± 5%, with a Bland-Altman mean difference of 0.3009% (limits: 0.2956 to 0.3063%). For PaO₂, the pre-alignment deviation was 3.9%, reduced to 0.4% post-alignment, both within the CLIA threshold of ± 10%, with a Bland-Altman mean difference of 4.0490 mmHg (limits: 3.9976 to 4.1004 mmHg). PCO₂ demonstrated a pre-alignment deviation of 4.2%, reduced to 0.19% post-alignment, both within the CLIA threshold of ± 10%, with a Bland-Altman mean difference of 0.3790 mmHg (limits: 0.3751 to 0.3829 mmHg). SvO₂ showed a pre-alignment deviation of 3%, which decreased to 0.8% post-alignment, both within the CLIA threshold of ± 5%, with a Bland-Altman mean difference of 0.7782% (limits: 0.7706 to 0.7858%). Finally, for SaO₂, the pre-alignment deviation was 2.6%, reduced to 0.1% post-alignment, both within the CLIA threshold of ± 5%, with a Bland-Altman mean difference of 0.9614% (limits: 0.9594 to 0.9634%). The Passing-Bablok regression analysis confirmed strong agreement, with slopes close to 1.0100 and intercepts near zero for all parameters. These results validate the QPS as a reliable and non-inferior tool for real-time blood gas monitoring during cardiopulmonary bypass, adhering to CLIA standards and ensuring clinical accuracy.
conclusionsThe findings support the accuracy of the Quantum Perfusion System compared to the BGA standard, demonstrating the system's capability to provide accurate, continuous blood gas monitoring during CPB. However, further studies are necessary to strengthen and confirm these results across broader and more varied clinical scenarios, for these reason as recommended by the manufacturers, the quantum monitoring system should only be used as a trending device.