Evidence map›Paper›PMID 40549303›Full record

ArticleEJNMMI physics2025

Establishing measurement traceability for quantitative SPECT imaging.

Andrew P Robinson, Kelley M Ferreira, Warda Heetun, Manuel Bardiès, Ana M Denis-Bacelar, Andrew J Fenwick, Michael Lassmann, Jill Tipping, Johannes Tran-Gia

Abstract read
In one paragraph

Article in EJNMMI physics, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Andrew P RobinsonNational Physical Laboratory, Hampton Road, London, TW11 0LW, UK. andrew.robinson@npl.co.uk.ORCID http://orcid.org/0009-0004-2552-5177
Kelley M FerreiraNational Physical Laboratory, Hampton Road, London, TW11 0LW, UK.
Warda HeetunNational Physical Laboratory, Hampton Road, London, TW11 0LW, UK.
Manuel BardièsCancer Research Institute of Montpellier, U1194, INSERM/ICM/Montpellier University, 208 Av des Apothicaires, 34298, Montpellier cedex 5, France.
Ana M Denis-BacelarNational Physical Laboratory, Hampton Road, London, TW11 0LW, UK.
Andrew J FenwickNational Physical Laboratory, Hampton Road, London, TW11 0LW, UK.
Michael LassmannDepartment of Nuclear Medicine, University Hospital Würzburg, Würzburg, Germany.
Jill TippingChristie Medical Physics and Engineering (CMPE), The Christie NHS Foundation Trust, Wilmslow Road, Manchester, M20 4BX, UK.
Johannes Tran-GiaDepartment of Nuclear Medicine, University Hospital Würzburg, Würzburg, Germany.

Funding

European Metrology Programme for Innovation and Research 19SIP01
6 · The paper itself

Abstract

backgroundSingle Photon Emission Computed Tomography (SPECT) is increasingly used as a quantitative modality, especially in the context of Molecular Radiotherapy, where the measurements are used as input to absorbed dose calculations for patient-specific dosimetry. Establishing measurement traceability is an essential step in providing confidence in quantitative measurements. This requires an unbroken chain of calibrations where uncertainties must be reported in all stages of calibration and for the final measurement result. Traceability ensures that a measurement result can be related to an underlying standard, allowing harmonisation of data, and facilitating comparison of results between sites.

methodsThe process of establishing measurement traceability for quantitative SPECT is demonstrated for the therapeutic radionuclide

resultsFor all activity measurements the dominant component in the activity uncertainty budget was the uncertainty on the radionuclide calibrator calibration factor, resulting in an average combined standard uncertainty of 1.57%. The resulting uncertainty on the SPECT Image Calibration Factor was 1.6%. An optional additional correction was included in the calibration to provide volume-based partial volume correction (PVC). Measurement traceability was extended for measurands using this additional correction. The activity recovery in the organ phantoms with PVC applied was 96(7)% for both the kidney and spleen.

conclusionsA manufacturer independent methodology for establishing measurement traceability for quantitative SPECT is demonstrated for

Indexed as

Measurement uncertaintyQuantitative imagingSPECTTraceability

Identifiers

PMID40549303
PMCPMC12185843

What Socratic holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390