Comparative Diagnostic Performance of the 36-Minute STANDARD M10 Fast Assay for Molecular Diagnosis of SARS-CoV-2, Influenza A and B, and Respiratory Syncytial Virus in Near-Patient Settings.
Article in Infectious diseases and therapy, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Diagnostic accuracy of rapid integrated molecular testing for major respiratory viral infections: a systematic review and meta-analysis.Frontiers in medicine · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
Abstract
introductionSARS-CoV-2, influenza A and B, and respiratory syncytial virus (RSV) are the major seasonal viruses that can cause severe illness. Rapid and accurate diagnosis of these viruses may improve patient management and surveillance efforts. Here, we assessed the diagnostic performance of the STANDARD M10 Flu/RSV/SARS-CoV-2 Fast (M10Fast) real-time polymerase chain reaction assay for near-patient diagnosis of the four viruses.
methodsThis retrospective validation study included 600 nasopharyngeal swab specimens previously tested using the standard-of-care laboratory-based Allplex Respiratory Panels 1-4 and Allplex SARS-CoV-2/FluA/FluB/RSV reference assays. Of these samples, 300 were positive for SARS-CoV-2, influenza A, influenza B, or RSV, while the remaining 300 samples were negative for the four viruses. Positive and negative percent agreements between the M10Fast index and reference tests were the primary endpoints. Additionally, analytical sensitivity of the M10Fast in terms of 95% limit of detection was estimated via serial dilutions of a positive reference material.
resultsOf 600 samples processed in the M10Fast, 590 (98.3%) were fully concordant. Positive percent agreement coefficients were 98.7% for influenza A and 100% for SARS-CoV-2, influenza B, and RSV. Negative percent agreement was 99.2% for influenza A, 99.4% for both SARS-CoV-2 and RSV, and 99.8% for influenza B. Discordant results were characterized by low viral loads with cycle threshold values of 38 or greater. The rate of invalid M10Fast runs was low (1.2%). Limit of detection of the M10Fast varied from 189 copies/mL for RSV to 541 copies/mL for the N gene of SARS-CoV-2.
conclusionsThe M10Fast, developed for near-patient settings, reliably detects SARS-CoV-2, FluA, FluB, and RSV in 36 min and its performance is comparable to standard laboratory-based assays.