ArticleDocumenta ophthalmologica. Advances in ophthalmology2026
The short wavelength electro-oculogram (SW-EOG) in best disease: preliminary results.
Article in Documenta ophthalmologica. Advances in ophthalmology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Review of the clinical electrooculogram - Part 2: the bestrophinopathies and modified protocols.Documenta ophthalmologica. Advances in ophthalmology · 2026Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThe clinical electrooculogram (EOG) is used for the diagnosis of bestrophinopathies to evaluate the function of the retinal pigment epithelium. The current ISCEV test protocol uses a broad band white light with luminance of 100 cd/m
methodsTo evaluate the clinical potential of the SW-EOG four genetically confirmed cases of Best Vitelliform Macular Dystrophy were tested using the standard white (100 cd/m
resultsIn all four cases the SW-EOG was reduced and provided an equivalent clinical measure for RPE dysfunction. All four participants rated the SW-EOG as being more comfortable than the white standard EOG test.
conclusionsThis is the first demonstration of an alternative stimulus for the EOG that provided clinically valid results with greater comfort than the current ISCEV protocol. Further studies are required to validate the SW-EOG as an alternative to the white broad band stimulus.