ArticleDocumenta ophthalmologica. Advances in ophthalmology2026

Intra- and inter-examiner reliability of ERGs recorded with two types of skin electrodes.

Ryunosuke Nagashima, Kumiko Kato, Keitaro Mizumoto, Sumine Mori, Daisuke Kurose, Iroha Fujimoto, Hisashi Matsubara, Hidetaka Kudo, Daphne L McCulloch, Mineo Kondo

Abstract read
In one paragraph

Article in Documenta ophthalmologica. Advances in ophthalmology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Ryunosuke NagashimaDepartment of Ophthalmology, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Kumiko KatoDepartment of Ophthalmology, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan. k-kato@clin.medic.mie-u.ac.jp.ORCID http://orcid.org/0000-0003-4172-3443
Keitaro MizumotoDepartment of Ophthalmology, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Sumine MoriDepartment of Ophthalmology, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Daisuke KuroseDepartment of Ophthalmology, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Iroha FujimotoDepartment of Ophthalmology, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Hisashi MatsubaraDepartment of Ophthalmology, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
Hidetaka KudoMayo Corporation, Inazawa, Japan.
Daphne L McCullochSchool of Optometry and Vision Science, University of Waterloo, Waterloo, Canada.
Mineo KondoDepartment of Ophthalmology, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo determine whether the intra-examiner and the inter-examiner reliabilities of the implicit times and amplitudes of the ERGs differed when recorded with two different types of skin electrodes. SUBJECTS AND

methodsEleven subjects (5 men and 6 women) were studied. The RETeval system was used to record ERGs, and the Sensor Strip of the RETeval system or the Red Dot electrocardiography electrodes (Red Dot) were used to record the ERGs. Flicker ERGs were recorded with a natural pupil. To test the intra-examiner reliability, one examiner positioned electrodes and recorded the ERGs three times on three different days from each subject. For the inter-examiner reliability, three different examiners recorded the ERGs from the 11 subjects. The implicit times and amplitudes of the flicker ERGs were analyzed using one-way intraclass correlation coefficients (ICC (1,1)) to assess intra-examiner variability and using a two-way intraclass correlation coefficient (ICC (2,1)) for inter-examiner variability.

resultsFor implicit times both the ICC (1,1) and ICC (2,1) were > 0.8 for both electrode types. For the amplitude, ICC (1,1) was > 0.8 for both electrodes, but the ICC (2,1) was lower at 0.474 for the Sensor Strip and 0.672 for the Red Dot electrodes. After Benjamini-Hochberg adjustments, the amplitudes did not differ between electrodes in either setting; implicit time was slightly shorter with the Red Dot than with Sensor Strip in the three-examiner setting (BH-adjusted P = 0.032; mean difference, 0.13 ms).

conclusionThe intra-examiner reliability was high for both types of electrodes, but the inter-examiner reliability indicated that the amplitudes of the ERGs tended to be less reliable for both types of electrodes. We conclude that standardizing the electrode placement and ensuring precise applications may help improve inter-examiner reproducibility, especially for the amplitude measurements.

Indexed as

ElectroretinographyRetinaAdultElectrodesFemaleHumansMaleMiddle AgedObserver VariationReproducibility of ResultsFlicker ERGReproducibilityRETevalSkin electrode

Identifiers

PMID42101571
PMCPMC13506558

What Socratic holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the Socratic graph.