Initial assessment of a species-specific rebound measurement mode for non-human primates.
Article in BMC veterinary research, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
Abstract
backgroundAccurate measurement of intraocular pressure in non-human primates is of significant importance in veterinary clinical practice, particularly for the prevention, detection, and management of glaucoma in non-human primates. However, no veterinary tonometer is currently designed specifically for non-human primates.
objectiveTo investigate the accuracy of the macaque-specific measurement mode of the iFalcon™ V100 rebound tonometer.
methodsIOP was increased stepwise from 5 mmHg to 90 mmHg in 3-5 mmHg increments. At each target pressure, the system was stabilized for 10 s, then the iFalcon™ V100 recorded six consecutive rebound readings and their average. Each step yielded 2-3 such average values. The results were compared with a high-precision invasive pressure sensor. Linear regression and Bland-Altman analyses were performed, and the measurement range was stratified (5-30, 30-60, 60-90 mmHg) for refined assessment.
resultsThe iFalcon™ V100 showed strong positive correlation with the invasive sensor (y = 0.99x + 0.42, R² = 0.99). The mean difference was - 0.036 mmHg (95% limits of agreement: -3.747 to 3.675 mmHg). Within the 5-30 mmHg range, agreement was excellent (mean difference: -0.063 mmHg). Acceptable agreement was maintained up to 90 mmHg.
conclusionsThis ex vivo study provides an initial assessment of the iFalcon™ V100 rebound tonometer in non-human primate eyes. The device showed good correlation and agreement with an invasive pressure sensor across a wide pressure range. While the sample size was limited, these preliminary findings support the potential of this species-specific measurement mode for IOP assessment in non-human primate research. Further studies with more eyes and in vivo conditions are needed to confirm its reliability and repeatability.