ArticleInternational journal of laboratory hematology2026
Rapid Centrifugation for Coagulation Testing: A Multianalyzer Validation Study Harmonizing Preanalytical Protocols for Total Laboratory Automation.
Article in International journal of laboratory hematology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesCentrifugation requirements differ between coagulation and clinical chemistry testing, potentially limiting workflow efficiency and total laboratory automation (TLA). Coagulation assays typically require standard centrifugation at 1500 g for 15 min, whereas chemistry samples use a rapid protocol at 2300 g for 5 min. This study evaluated the impact of rapid centrifugation on prothrombin time (PT), activated partial thromboplastin time (APTT), and D-dimer results.
methodsBlood samples were prospectively collected from volunteers in 3, 2, and pediatric 1 mL citrate tubes and were processed using standard or rapid centrifugation. Testing was performed for PT, APTT, and D-Dimer on Sysmex CN3000 and ACL-TOP 750 analyzers, with residual platelet counts assessed.
resultsA total of 130 samples were analyzed (41 from 3 mL tubes, 41 from 2 mL tubes, and 39 from 1 mL tubes). Rapid centrifugation produced significantly higher residual platelet counts than the standard protocol (39.2 × 10
conclusionsRapid centrifugation at 2300 g for 5 min provides coagulation results comparable to the standard protocol and may facilitate integration of coagulation testing into TLA workflows.