ArticleCardiovascular engineering and technology2026
Reducing Stent-Induced Blooming Artifacts Using Virtual Monoenergetic Imaging Reconstructions in a Superficial Femoral Artery Phantom: An In-Vitro Study.
Article in Cardiovascular engineering and technology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeSpectral computed tomography derived virtual monoenergetic imaging (VMI) reconstructions have been shown to reduce stent-induced blooming artifacts thus minimizing in-stent lumen underestimation. We have investigated the use of VMI reconstructions to reduce the blooming artifact for stent markers, single and double stent configurations.
methodsTwo partially overlapping covered stents were deployed in a superficial femoral artery phantom. A conventional and twelve VMI reconstructions (40-100 keV and 120-200 keV with 10 and 20 keV increments, respectively) were created. Contrast-to-noise ratio, signal-to-noise ratio, attenuation difference between native and in-stent vessel lumen and full width at half maximum (FWHM) of the stent peaks were evaluated. The in-stent lumen diameter was measured, and the stent evaluated using a 5-point scale by three interventional radiologists.
resultsThe 80-100 keV VMI reconstructions for the single stent configuration and the 140 keV VMI reconstruction for the other configurations resulted in the smallest lumen underestimation and best quality score on blooming. The underestimation was reduced from 21.5-29.1% for the conventional reconstruction to 14.1-18.6% for the optimal VMI reconstructions. No significant differences were observed in FWHM between the conventional reconstruction and the 50-200 keV VMI reconstructions.
conclusionsVMI was found to minimize stent-induced blooming artifacts and improve in-stent lumen assessment. More stent material requires higher-energy VMI reconstructions to reduce blooming. The absence of significant differences in quantitative parameters suggests that the number of pixels affected by the blooming artifact is not reduced using VMI reconstructions.